Precise power for every X-ray room. The ISAIND HF series brings together five high-frequency generators from 40 to 150 kV and up to 800 mA, ready for new rooms and for upgrading analog rooms to digital radiography (DR). They integrate with your tube, your bucky and your flat panel detector, and are calibrated from software.

Two ISAIND HF high-frequency X-ray generator cabinets
ISAIND HF · High-frequency X-ray generators

No published price and no online checkout. Configuration and availability are confirmed when quoting.

Key figures

  • 40–150 kV · kV range
  • ≤1 ms · kV rise @ 100 kV
  • 10–800 mA · Tube current
  • 32–56 kW · Nominal power · up to 70 kW

Models

  • HF-32S · single phase 220 VAC · 32 / 40 kW
  • HF-40S · single phase 220 VAC · 40 / 50 kW
  • HF-40T · three phase 380 VAC · 40 / 50 kW
  • HF-50T · three phase 380 VAC · 50 / 63 kW
  • HF-56T · three phase 380/400 VAC · 56 / 70 kW

Downloads

WhatsApp +593 98 998 5755 · Jardines del Salado Mz 207 V15, Guayaquil, Ecuador · info@isaind.com · isaind.com

Accurate exposures, simple integration and a generator built to last

  • Ready for DR retrofit

    Flat panel detector sync interlock, Bucky Start/Ready signals and a communication protocol for imaging software: convert analog rooms to digital without replacing the whole room.

  • Works with your tube

    Compatible with single-phase, three-phase, low-speed, high-speed and dual-speed tubes. Dual-speed automatic rotor control with variable-frequency drive.

  • kV and mA precision

    kV ±5% @ ≥100 kV, tube current ±(5% + 1 mA) and mAs ±(5% + 0.2 mAs) @ ≥10 ms. Built-in tube curves and automatic filament calibration.

  • Fast response

    kV rise ≤1 ms @ 100 kV and exposures from 1 to 10,000 ms, driven by a fast-response high-voltage inverter control algorithm.

  • Rugged high-voltage tank

    Monolithic die-cast tank, vacuum oil filling and high insulation level: withstands 240 kV. Five-stage EMC filtering.

  • Easy integration and calibration

    Mini or full console plus configuration software: filament calibration, profile import, log export and interface status setup.

ISAIND HF + ISAIND Fluoro bundle

ISAIND HF generators are also offered as a bundle together with ISAIND Fluoro, our imaging software: generator and software from a single supplier, with unified integration and support. The scope of each bundle is defined with ISAIND for your project; configuration and compatibility are confirmed on site, with no universal compatibility declared.

  • A single supplier

    The X-ray generator and the ISAIND Fluoro software come from the same company, with one point of contact for your project.

  • Unified integration

    ISAIND coordinates the integration of the generator and the software with your system and verifies the configuration on site.

  • Unified support

    Technical support, training and maintenance handled by the same team, directly via WhatsApp and email.

Who is the ISAIND HF series for?

  • Integrators & manufacturers

    A generator with standard interfaces to equip conventional and DR rooms with your own table, bucky and detector.

  • Retrofit workshops

    Replace aging generators and move analog rooms to digital with flat panel detector synchronization.

  • Distributors

    Five single- and three-phase models to cover everything from clinics to hospitals.

  • Hospitals & clinics

    Renew your room’s generator and gain precision and repeatability in every exposure.

Technical specifications

By model

ModelHF-32SHF-40SHF-40THF-50THF-56T
PhaseSingle phaseSingle phaseThree phaseThree phaseThree phase
Input voltage220 VAC · 50/60 Hz220 VAC · 50/60 Hz380 VAC · 50/60 Hz380 VAC · 50/60 Hz380/400 VAC · 50/60 Hz
Nominal / maximum power32 kW / 40 kW40 kW / 50 kW40 kW / 50 kW50 kW / 63 kW56 kW / 70 kW
Tube current range10–400 mA10–500 mA10–500 mA10–630 mA10–800 mA
mAs range0.1–400 mAs0.1–400 mAs0.1–1000 mAs0.1–1000 mAs0.1–1000 mAs
Weight110 kg110 kg78 kg78 kg78 kg

Common to all models

kV range40–150 kV
kV accuracy±5% @ ≥100 kV · 4 kV @ <100 kV
kV rise time≤1 ms @ 100 kV (10%–90%)
Tube current accuracy±(5% + 1 mA) @ ≥10 ms · ±(20% + 1 mA) @ <10 ms
mAs accuracy±(5% + 0.2 mAs) @ ≥10 ms · ±(10% + 0.2 mAs) @ <10 ms
Exposure time1–10,000 ms
Exposure time accuracy±(2% + 0.5 ms) @ ≥10 ms · ±(5% + 1 ms) @ <10 ms
Rotor controlDual-speed automatic control
OptionalAEC / DAP
Dimensions700 × 400 × 680 mm
High-voltage tankMonolithic die-cast, vacuum oil filling, 240 kV withstand voltage

Interfaces

Room lightRelay-type output signal for indicator light
Bucky StartOutput signal for X-ray exposure synchronization
Bucky ReadyInput signal for X-ray exposure synchronization
Tube thermal switchInput signal for X-ray tube thermal switch
InterlockInput for flat panel detector exposure synchronization
Door interlockInput detecting the status of the shielded lead door
Bucky supply24 VDC / 110 VAC / 220 VAC output to Bucky
Collimator supply24 V output to the collimator
AECIonization chamber connection for automatic exposure control
DAPSoftware-based dose area product calculation
Emergency stopExternal emergency stop button connection
Communication protocolCommunication with imaging software and custom protocol
X-ray tubeCompatible with single-phase, three-phase, low-speed, high-speed and dual-speed tubes

Specifications subject to change without notice; please confirm the configuration with ISAIND before purchase. Images for reference only. Import and registration requirements vary by country.

A technical partner for your project

ISAIND specializes in medical imaging: we refurbish, remanufacture, recondition and maintain X-ray equipment and C-arms. With the ISAIND HF series we help you choose the model, integrate it into your room and support you for its entire service life.

  • Choosing the model

    Tell us your tube, your mains supply (220 or 380/400 VAC) and your detector, and we will recommend the right model.

  • Installation and calibration

    Setup, filament calibration and functional testing with your equipment.

  • Technical support

    Engineers who know the system in depth, with direct support via WhatsApp and email.

  • Remanufacturing experience

    Real knowledge of generators, tubes and detectors from our refabrication and reconditioning work.

Request the full datasheet and a proposal for your project

For integrators, retrofit workshops, distributors and hospitals worldwide. Tell us what you need and we will reply with clear technical information and no commitment.

Technical service and parts

For this equipment you can also request technical service or quote parts.